Hamilton Medical Infant Flow Sensor recall
Hamilton Medical Infant Flow Sensor, single use, (3.1m), Part number 260179. The infant flow sensor is an accessory for the Hamilton Medical…
Why was Hamilton Medical Infant Flow Sensor recalled?
An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
PN 260179, all lot codes.
- Product
- Hamilton Medical Infant Flow Sensor, single use, (3.1m), Part number 260179. The infant flow sensor is an accessory for the Hamilton Medical…
- Recalling firm
- Hamilton Medical, Inc.
- Quantity
- 154 total, all part numbers
- Distribution
- US Nationwide Distribution.
- Recall date
- July 8, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)