System 1 Base recall
System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for…
Why was System 1 Base recalled?
Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog number: 801763 and serial numbers: 0011-0335, 0337-0610, 1001-1034, 1037-1049, 1100, 1103-1106, 1109-1445, and 1447-1450.
- Product
- System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for…
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Quantity
- approx. 993 units
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV. Internationally to: Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam,
- Recall date
- July 18, 2012
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)