FDAClass IIJune 18, 2025

2.5 V replacement rechargeable batteries for Hillrom recall

2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600

Why was 2.5 V replacement rechargeable batteries for Hillrom recalled?

The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI 00732094139488, Lot Number 25-043, Exp. August 2027

Product
2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
Recalling firm
Baxter Healthcare Corporation
Quantity
277 units
Distribution
US Nationwide distribution.
Recall date
June 18, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls