FDAClass IIJuly 17, 2019

Radrex-i X-Ray System recall: Software Issue

Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system…

Why was Radrex-i X-Ray System recalled?

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

No./Serial No./Manufacture Date: No 1./U8A1312001/JANUARY 2013; No.2/U8A1312002/JANUARY 2013; No.3/U8A1342003/APRIL 2013; No.4/U8A1362004 /JUNE 2013; No. 5 U8B13X2006/OCTOBER 2013; No. 6 U8B13Z2008/DECEMBER 2013; No. 7 U8B1412009/JANUARY 2014; No. 8 U8B1432010 /MARCH 2014; No. 9 U8B1442011/APRIL 2014; No.10 U8B1462013/JUNE 2014; No.11 U8B1472014/JULY 2014; No.12 U8B1482015/AUGUST 2014; No.13 U8B14Y2016/NOVEMBER 2014; No.14 U8B14Z2017/DECEMBER 2014; No.15 U8B14Z2018/DECEMBER 2014; No.16 U8B14Z2019/DECEMBER 2014; No.17 U8B1512020/JANUARY 2015; No.18 U8B1522021/FEBRUARY 2015; No.19 U8B1522022/FEBRUARY 2015; No.20 U8B1522023/FEBRUARY 2015; No.21 U8B1522024/FEBRUARY 2015; No.22 U8B1612025/JANUARY 2016; No.23 U8B1632026/MARCH 2016; No.24 U8B1642027/APRIL 2016; No.25 U8B1642028/APRIL 2016; No.26 U8B1642029/APRIL 2016; No.27 U8B1662030 /JUNE 2016; No.28 U8B1677001/JULY 2016; No.29 U8B16X2032/OCTOBER 2016; No.30 U8B1722033 /FEBRUARY 2017; No.31 U8B1752034/MAY 2017; No.32 U8B1762035/JUNE 2017; No.33 U8B1782036 /AUGUST 2017; No.34 U8B17Z2037 /DECEMBER 2017; No.35 U8C1812038/JANUARY 2018; No.36 USA1462001/JUNE 2014; No.37 USA1492002/SEPTEMBER 2014; No.38 USA14Z2003/DECEMBER 2014; No.39 USA1512004/JANUARY 2015; No.40 USA1532005/MARCH 2015; No.41 USA15Z2006/DECEMBER 2015; No.42 USA1662007/JUNE 2016; No.43 USA1682008/AUGUST 2016; No.44 USA16X2009/OCTOBER 2016; No.45 USA16Y2011/NOVEMBER 2016; No.46 U8A1372005/JULY 2013; No.47 U8B13X2007/OCTOBER 2013; No.48 USA16X2010/OCTOBER 2016

Product
Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system…
Recalling firm
Canon Medical System, USA, INC.
Quantity
46
Distribution
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
Recall date
July 17, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)