Siemens Cios Alpha VA20-mobile X-Ray system Material # recall
Siemens Cios Alpha VA20-mobile X-Ray system Material # 10308191 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray…
Why was Siemens Cios Alpha VA20-mobile X-Ray system Material # recalled?
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 12425, 109, 10063, 12946
- Product
- Siemens Cios Alpha VA20-mobile X-Ray system Material # 10308191 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 4 units
- Distribution
- US Nationwide distributions.
- Recall date
- May 20, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)