Access 2 Immunoassay System recall
Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or…
Why was Access 2 Immunoassay System recalled?
Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial No. 505143, 504805, 501368, 504231
- Product
- Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 4 units total (1 unit in US)
- Distribution
- US, Morocco, Venezuela
- Recall date
- June 22, 2016
- Classification
- Class II
- Status
- Terminated