FDAClass IJuly 7, 2021

LeadCare Plus Blood Lead Test Kit recall: Lead

LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004

Why was LeadCare Plus Blood Lead Test Kit recalled?

Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers/Expiration Date: 2011MU - 25MAR22. 1st expansion 6/21/2021: 2104MU - 25AUG22; 2108MU - 31MAR22.

Product
LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004
Recalling firm
Magellan Diagnostics, Inc.
Quantity
320 kits (96 tests/kit=30,720 tests)
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.
Recall date
July 7, 2021
Classification
Class I
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls