Brand Name: HiResolution" Bionic Ear System Product recall
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130…
Why was Brand Name: HiResolution" Bionic Ear System Product recalled?
The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Code: All lots manufactured up to February 3, 2026. Refer to Exhibit_0006
- Product
- Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130…
- Recalling firm
- Advanced Bionics, LLC
- Quantity
- 45173
- Distribution
- Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, China, Chile, Croatia, Cuba, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad and Tobago, United Arab Emirates, and United Kingdom.
- Recall date
- May 6, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)