AIDBAGs are first aid kits of convenience composed of recall: Software Issue
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate…
Why was AIDBAGs are first aid kits of convenience composed of recalled?
Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
- Product
- AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate…
- Recalling firm
- North American Rescue LLC.
- Quantity
- 21
- Distribution
- US: CA, WA
- Recall date
- May 6, 2026
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)