iChem¿VELOCITY Urine Chemistry System recall
iChem¿VELOCITY Urine Chemistry System-
Why was iChem¿VELOCITY Urine Chemistry System recalled?
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog #700-7176-001 and #700-7177-001, All part numbers
- Product
- iChem¿VELOCITY Urine Chemistry System-
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 1557 units
- Distribution
- US Distribution: Nationwide Foreign Distribution: Worldwide
- Recall date
- May 20, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)