Femoral Nail recall
Femoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Number: 1851-0830S.
Why was Femoral Nail recalled?
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog Number: 1851-0830S. GTIN: 04546540647184. Batch lot number: K09F17F
- Product
- Femoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Number: 1851-0830S.
- Recalling firm
- Stryker GmbH
- Quantity
- 2
- Distribution
- Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands.
- Recall date
- June 18, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)