Onetouch Delica Lancets recall
ONETOUCH DELICA LANCETS 33G 100 COUNT Item Number: 5541628
Why was Onetouch Delica Lancets recalled?
Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
NDC/UPC: 812608030019 All lots numbers shipped from 2/23/201 to 3/10/2021
- Product
- ONETOUCH DELICA LANCETS 33G 100 COUNT Item Number: 5541628
- Recalling firm
- Cardinal Health Inc.
- Quantity
- 4,896 boxes of 100 lancets
- Distribution
- US Nationwide distribution in the states of FL, GA, SC.
- Recall date
- June 30, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)