FDAClass IIJuly 11, 2012

Vacuette Quickshield Complete Plus recall

Vacuette Quickshield Complete Plus, 21G x 1 (0.8 x 25 mm), Greiner bio-one. The Quick shield Complete Plus is intended to be used only with Vacuette…

Why was Vacuette Quickshield Complete Plus recalled?

: The Quickshield Complete Plus may have a defect (split at the base), which can cause the green safety shield attachment of the Quickshield holder to be loose and unstable.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item #450239, Lot #G110601I and Item #450235, Lot #G110302J.

Product
Vacuette Quickshield Complete Plus, 21G x 1 (0.8 x 25 mm), Greiner bio-one. The Quick shield Complete Plus is intended to be used only with Vacuette…
Recalling firm
Greiner Bio-One North America, Inc.
Quantity
19.5 cases (5,850 pieces) and 244 cases (73,200 pieces)= 263.5 (73258 pieces total)
Distribution
Nationwide Distribution including IL, MS, OH, PA, TN and VA.
Recall date
July 11, 2012
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls