Freestyle Lancets recall
FREESTYLE LANCETS 28G 100CT Item Number:2975670
Why was Freestyle Lancets recalled?
Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
NDC/UPC: 699073130013 All lots numbers shipped from 2/23/201 to 3/10/2021
- Product
- FREESTYLE LANCETS 28G 100CT Item Number:2975670
- Recalling firm
- Cardinal Health Inc.
- Quantity
- 377 boxes of 100 lancets
- Distribution
- US Nationwide distribution in the states of FL, GA, SC.
- Recall date
- June 30, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)