FDAClass IJune 6, 2018

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump recall

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 &…

Why was CARDIOSAVE Hybrid Intra-Aortic Balloon Pump recalled?

Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Serial Numbers distributed from 03/06/2012 to 04/26/2018

Product
CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 &…
Recalling firm
Maquet Datascope Corp - Cardiac Assist Division
Quantity
4300
Distribution
The devices were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Guam. The devices were distributed to the following foreign countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, Chile, China, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrica, Taiwan, Thailand, Turkey, UAE, UK, Vietnam, Yemen.
Recall date
June 6, 2018
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls