LDL-Cholesterol gen.2. The Boehringer Mannheim Direct recall
LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein…
Why was LDL-Cholesterol gen.2. The Boehringer Mannheim Direct recalled?
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog numbers 04714423190; 04711220190; 05230438190; 05230446190.
- Product
- LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein…
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 6255
- Distribution
- US (nationwide) including PR.
- Recall date
- July 8, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)