FDAClass IIMay 30, 2018

Prismaflex 7.XX ROW recall

Prismaflex 7.XX ROW, Product Code 114870 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for…

Why was Prismaflex 7.XX ROW recalled?

Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Lots and Serial Numbers

Product
Prismaflex 7.XX ROW, Product Code 114870 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for…
Recalling firm
Baxter Healthcare Corporation
Quantity
2772 units
Distribution
Worldwide distribution: United States (including Puerto Rico), and countries of: Canada, Australia, New Zealand, China, Bangladesh, Japan, Malaysia, Myanmar, Sri Lanka, India, Singapore, Vietnam, Hong Kong, Korea, Taiwan, Indonesia, Bruni, Philippines, Thailand, Argentina, Chile, Costa Rica, Dominican Republic, Haiti, Barbados, Bermuda, Venezuela, Cuba, Mexico, Colombia, Brazil and EMEA.
Recall date
May 30, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls