FDAClass IIMay 10, 2017

Schiller AG ISA CO2 Sensor recall

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the…

Why was Schiller AG ISA CO2 Sensor recalled?

The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841.

Product
Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the…
Recalling firm
Merge Healthcare, Inc.
Quantity
23 sensors (PhaseIn Units)
Distribution
US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
Recall date
May 10, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)