FDAClass IIJuly 8, 2015

Creatinine Plus. Intended for use on Hitachi/MODULAR recall

Creatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative determination of the creatinine concentration in serum, plasma, and…

Why was Creatinine Plus. Intended for use on Hitachi/MODULAR recalled?

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog Numbers: 11775685216; 11875566216; 11875582216.

Product
Creatinine Plus. Intended for use on Hitachi/MODULAR systems for the quantitative determination of the creatinine concentration in serum, plasma, and…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
6,255 analytical units distributed
Distribution
US (nationwide) including PR.
Recall date
July 8, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls