Smith&nephew Anthology Porous High Offset Stem recall
smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.
Why was Smith&nephew Anthology Porous High Offset Stem recalled?
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 17LM22019, 17LM10934, 17LM05900
- Product
- smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 13 units
- Distribution
- US nationwide distribution.
- Recall date
- May 30, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)