FDAClass IIJune 15, 2016

Liko Universal SlingBars Universal SlingBar" 350 recall

Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal…

Why was Liko Universal SlingBars Universal SlingBar" 350 recalled?

The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk for breakage with the result of a free fall of the patient. This hazard could cause potentially minor to catastrophic injuries to a patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Slingbars that are affected by this field action (in the table below) have been distributed within the following serial number range: 1200101 to 1370151 Model Product/Model # Universal SlingBar" 350, 3156074 Universal SlingBar" 350 QRH, 3156084 Universal SlingBar" 350 R2R, 3156094 Universal SlingBar" 450, 3156075 Universal SlingBar" 450 QRH, 3156085 Universal SlingBar" 450 R2R, 3156095 Universal SlingBar" 600 3156076 Universal SlingBar" 600 QRH 3156086

Product
Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal…
Recalling firm
Hill-Rom, Inc.
Quantity
163,561
Distribution
Worldwide and US Nationwide Distribution AE, AO, AR , AT , AU , AZ , BD , BE , BG , BH , BM , BN , BR , CA , CH , CL , CN , CO , CS , CZ , DE , DK , EG , ES , FI , FR , GB , GR , HK , HR , HU , ID , IE , IL , IN , IR , IS , IT , JO , JP , KR , KW , LB , LI , LK , LT , LU , LV , LY , MA , MT , MX , MY , NL , NO , NZ , OM , PH , PK , PL , PT , QA , RO , RS , RU, SA , SE , SG , SI , SK , SY , TH , TN , TR , TW , UA , US , UY and ZA.
Recall date
June 15, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls