FDAClass IIJune 15, 2016

DePuy Synthes 8.0 mm Flexible Shaft Reamer recall

DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43

Why was DePuy Synthes 8.0 mm Flexible Shaft Reamer recalled?

Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995.

Product
DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
Recalling firm
Synthes (USA) Products LLC
Quantity
244 units
Distribution
US nationwide distribution.
Recall date
June 15, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls