Total Knee Arthroplasty recall: Software Issue
Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako…
Why was Total Knee Arthroplasty recalled?
Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: (01)7613327566444(10). Software: TKA 2.0 + TKA 1.0.1 + THA 4.1 + PKA 3.0.2 + MGO 1.2.2. Application Part Number: 700001590414. Revision: AA, AB, & AC.
- Product
- Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako…
- Recalling firm
- Stryker Orthopaedics
- Quantity
- 1,635 units
- Distribution
- Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, Thailand, Turkey, UAE, and UK
- Recall date
- June 5, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)