DAC 044 115cm recall: Endotoxin
DAC 044 115cm, Catalog: 90760
Why was DAC 044 115cm recalled?
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog: 90760, UDI-DI: 07613327313918, Lot: 156628
- Product
- DAC 044 115cm, Catalog: 90760
- Recalling firm
- Stryker Neurovascular
- Quantity
- 81
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Recall date
- June 5, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)