FDAClass IIJune 5, 2024

Trevo Pro 14 Microcatheter recall: Endotoxin

Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238

Why was Trevo Pro 14 Microcatheter recalled?

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog/UDI-DI/Lot: 90231/00815742002317/121050, 129849, 129882, 130285, 136439, 136628, 143375, 143476; 90238/00815742002386/136009, 198334

Product
Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238
Recalling firm
Stryker Neurovascular
Quantity
648
Distribution
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
Recall date
June 5, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls