FDAClass IJune 6, 2018

Heartware Medtronic HVAD System for cardiac use. recall

Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403…

Why was Heartware Medtronic HVAD System for cardiac use. recalled?

Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ALL SERIAL NUMBERS UPN/GTIN: (a) 1400 and 1401: distributed prior to GTN requirement 1403: 00888707000116, 00888707000475 1407: '00888707000291, '00888707000727, '00888707000710, 00888707000734, '00888707000741, '00888707000765, '00888707000758, '00888707000789, '00888707000772, '00888707000499, '00888707000796, '00888707000802, '00888707000819, '00888707000482, '00888707004534, '0888707001670, '00888707001656, '00888707001663, '00888707001670, '00888707001687, '00888707001694, '00888707001700, '00888707001717, '00888707001724, '00888707001731, '00888707001748, '00888707001755, '00888707001762, '00888707002813, '00888707002820, '00888707002837, '00888707002844, '00888707002851, '00888707001472 1420: '00888707000420, '00888707000437, '00888707002660 (b) 1435: '00888707000109 1440: '00888707000260, '00888707001885, '00888707001496, '00888707002745 (c) 1425: '00888707000093 1430: '00888707000307, '00888707000826, '00888707000833, '00888707000840, '00888707000857, '00888707000871, '00888707000864, '00888707000888, '00888707000901, '00888707000505, '00888707000895, '00888707000918, '00888707001489, '00888707001779, '00888707001786, '00888707001793, '00888707001809, '00888707001816, '00888707001823, '00888707001830, '00888707001847, '00888707001854, '00888707001861, '00888707001878, '00888707002738, '00888707002769, '00888707002776, '00888707002783, '00888707002790, '00888707002806, '00888707004817 (d) 1650DE: 00888707000369, 00888707001373, '00888707001588, '00888707001366, '00888707002684 '00888707002646, '00888707002653, '00888707000376

Product
Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits, Product numbers: 1400, 1401, 1403…
Recalling firm
Heartware
Quantity
175878 devices
Distribution
Worldwide Distribution including US Nationwide, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea, , Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, and Vietnam.
Recall date
June 6, 2018
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls