Delfi Medical Innovations recall
Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile
Why was Delfi Medical Innovations recalled?
Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
18J17
- Product
- Delfi Medical Innovations, VariFit Contour Cuffs, single use adult thigh cuff, single port, single bladder, REF 9-9300-003. Rx Only, Sterile
- Recalling firm
- Delfi Medical Innovations, Inc.
- Quantity
- 59 units
- Distribution
- US: Ohio OUS: Austrailia
- Recall date
- May 13, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)