FDAClass IIJune 5, 2024

FDR Visionary Suite - Intended to generate digital or recall

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and…

Why was FDR Visionary Suite - Intended to generate digital or recalled?

Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002

Product
FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and…
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
18 units
Distribution
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Recall date
June 5, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls