Signature Laparoscopic Instruments recall
Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862100.
Why was Signature Laparoscopic Instruments recalled?
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/REF Number: 5862100. UDI-DI: B33158621000. Lot Number: M28183. Serial Numbers: 1932, 1933, 1934, 1935, 1936, 1937, 1938, 1939.
- Product
- Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862100.
- Recalling firm
- Silex Medical, LLC
- Quantity
- 8 units
- Distribution
- US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.
- Recall date
- June 11, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)