Vercise Genus Deep Brain Stimulation recall
Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
Why was Vercise Genus Deep Brain Stimulation recalled?
Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026
- Product
- Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 14,977
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.
- Recall date
- June 5, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)