daVinci XI Sureform 60 reload recall
daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Product Usage: are intended to be…
Why was daVinci XI Sureform 60 reload recalled?
The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot numbers: T10180511; T10180718; T10180821;T10180822; T10180906; T10180913; T10180925; T10181008; T10190204; T10190208; T10190213; T11180816; T11190219
- Product
- daVinci XI Sureform 60 reload, 60 mm Green 6- Row, 4.6 mm, REF 48360G-08, UDI: 00886874113844, Rx Only, Sterile EO - Product Usage: are intended to be…
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 22,752 reloads
- Distribution
- Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.
- Recall date
- May 13, 2020
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)