FDAClass IIMay 23, 2018

5mm Endopath Xcel with Optiview Technology recall

5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 150mm shaft length, 2B5XT

Why was 5mm Endopath Xcel with Optiview Technology recalled?

Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lots P4TD1E R4001U R4052P

Product
5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 150mm shaft length, 2B5XT
Recalling firm
Ethicon Endo-Surgery Inc
Quantity
1500
Distribution
Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.
Recall date
May 23, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)