8000-021-002 recall: Software Issue
8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional…
Why was 8000-021-002 recalled?
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Sofware Version: NOVA 3.6.0 RC16, Product Code: 8000-021-002; UDI/DI: 07613327514179; Serial Numbers: 10984, 10110, 10164, 10454, 10762, 10387, 10120, 10252, 10414, 10612, 10408, 10673, 11013, 11032, 10566, 10974, 0310-B, 10326, 10442, 10729, 10681, 10552, 10088, 10476, 10307, 10121, 10371, 0252-B, 10262, 10602, 10613, 10381, 10234, 200774, 10135, 0304-B, 10303, 10257, 10639, 10445, 10684, 10740, 10986, 10413, 10079, 10528, 10565, 10482, 10209, 10574, 10606, 10229, 10214, 10797, 0335-B, 10284, 10465, 10295, 10308, 0282-B, 83B750AD, 0287-B, 10763, 10434, 11025, 10578, 10473, 10596, 0291-B, 10624, 10591, 11092, 10758, 10097, 10294, 0251-B, 10597, 10769, 0309-B, 10765, 0313-B, 0281-B, 10513, 11088, 10057, 10337, 10464, 10621, 10699, 10509, 10421, 10549, 10003, 10499, 10661, 10668, 0292-B, 10416, 0263-B, 10654, 10571, 0329-B, 10934, 10798, 11022, 10610, 10766, 10471, 10437, 10648, 10410, 10403, 10159, 10192, 10372, 10189, 10193, 10642, 10390, 10771, 10202, 10806, 10237, 10296, 0244-B, 0307-B, 10034, 10440, 10704, 10101, 10153, 10098, 10388, 10720, 10943, 10944, 10994, 10695, 10608, 10615, 10054, 10076, 10015, 10089, 10208, 10776, 10780, 11111, 10586, 10759, 10431, 10772, 10376, 10563, 10978, 10989, 10369, 10641, 10796, 10558, 10660, 10071, 10987, 10953, 10982, 10744, 10026, 0325-B, 10959, 11000, 10453, 0332-B, 10055, 10335, 10742, 10649, 10983, 10512, 10958, 10268, 10640, 0280-B, 11075, 11117, 10356, 10429, 10432, 10659, 10677, 10319, 10311, 10332, 10757, 10809, 10523, 10382, 10451, 10495, 10997, 11116, 11018, 10083, 10672, 10775, 10053, 10195, 10330, 10161, 10007, 10328, 10949, 11031, 10781, 10990, 10350, 10750, 10191, 10033, 10056, 10446, 10078, 10516, 10373, 10722, 10136, 10210, 10091, 10255, 10253, 10060, 10427, 10162, 10297, 10731, 10935, 0285-B, 10336, 10760, 10386, 10347, 10286, 10165, 10041, 10456, 10312, 10417, 10500, 10133, 10224, 10031, 10040, 10100, 10784, 10248, 10423, 10215, 10951, 10092, 0260-B, 10185, 10127, 10112, 10790, 10299, 10316, 10435, 10459, 10075, 10479, 10357, 10627, 10732, 10734, 0343-B, 0320-B, 10517, 10455, 10590, 11030, 10383, 10043, 10096, 10789, 10631, 10489, 10791, 10220, 10658, 10348, 10143, 10114, 10625, 10799, 11108, 10103, 10522, 10126, 10066, 10980, 11102, 0318-B, 11005, 10177, 10662, 10166, 10197, 10300, 10406, 10181, 0326-B, 10443, 10331, 10364, 10368, 10592, 10062, 10366, 10349, 10353, 11097, 10155, 10218, 10227, 10230, 10315, 10618, 10021, 10787, 11038, 0271-B, 10365, 10637, 10141, 10525, 10341, 10433, 10696, 0333-B, 10401, 10633, 10211, 10231, 10052, 10794, 10685, 10157, 10213, 10081, 10123, 10090, 10199, 10678, 11020, 10216, 10377, 10424, 10212, 10361, 10059, 10109, 10736, 10344, 10693, 10977, 11004, 10016, 10168, 10171, 10691, 10006, 10458, 10324, 10156, 10190, 10221, 10363, 10475, 0322-B, 0339-B, 10692, 10679, 10082, 10573, 10603, 10267, 10541, 10999, 10169, 10628, 10576, 10345, 10738, 10354, 10238, 0259-B, 10077, 0306-B, 10461, 10172, 10243, 10233, 10327, 0327-B, 11023, 11114, 10450, 10398, 10404, 0303-B, 10400, 10412, 10593, 10072, 10514, 10239, 10069, 10244, 10249, 10761, 10359, 10739, 10352, 10163, 10374, 10698, 10342, 10607, 10635, 10058, 10425, 10419, 10950, 10405, 10422, 10223, 0308-B, 10010, 10594, 10205, 10370, 10474, 10247, 10020, 10174, 10184, 10188, 10128, 11003, 10689, 10701, 10795, 10074, 10266, 10288, 10008, 10203, 10966, 10438, 10955, 10073, 10080, 10333, 10777, 10508, 10325, 10086, 10104, 10469, 10755, 10810, 10671, 10152, 10393, 10657, 10105, 10298, 10524, 10805, 10375, 10402, 10094, 10358, 10498, 10501, 10384, 10391, 10993, 10064, 10175, 10397, 10560, 10570, 10630, 10132, 10579, 10439, 10415, 10306, 10343, 10477, 10323, 10737, 10589, 10304, 10338, 10460, 10598, 10779, 10988, 10346, 11082, 10436, 10362, 10334, 10682, 10653, 10583, 10724, 10725, 10407, 10380, 10409, 10595, 10137, 10656, 10389, 10611, 10638, 11036, 11037, 10577, 10609, 10587, 10626, 10965, 11017, 11112, 10547, 10954, 11029, 11026, 10979, 10985, 11080, 10981, 11099, 10207, 10049, 10580, 10144, 10723.
- Product
- 8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional…
- Recalling firm
- Stryker Leibinger GmbH & Co. KG
- Quantity
- 705 units
- Distribution
- US Nationwide distribution.
- Recall date
- May 29, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency