FDAClass IIJune 8, 2016

Torque Limiting Handle: PRECIMED recall: Sterility Concern

Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20…

Why was Torque Limiting Handle: PRECIMED recalled?

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

6307760001

Product
Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20…
Recalling firm
Greatbatch Medical
Quantity
2
Distribution
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Recall date
June 8, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls