Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit recall
Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
Why was Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit recalled?
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a) REF IAB-05830-U, UDI: (01)10801902172850(17)240531(10)18F22F0023, (01)10801902172850(17)240630(10)18F22G0005, (01)10801902172850(17)240630(10)18F22G0010, (01)10801902172850(17)240731(10)18F22H0033, (01)10801902172850(17)250831(10)18F23J0031, (01)10801902172850(17)250930(10)18F23K0036, (01)10801902172850(17)250930(10)18F23K0042, (01)10801902172850(17)251130(10)18F23M0001, (01)10801902182019(17)250430(10)18F23E0032, (01)10801902182019(17)250831(10)18F23J0002, (01)10801902182019(17)250831(10)18F23J0082, (01)10801902182019(17)250930(10)18F23K0025, (01)10801902182019(17)260331(10)18F24C0131; b) REF IAB-05840-U, UDI: (01)10801902161922(17)240630(10)18F22G0051, (01)10801902161922(17)241231(10)18F23A0002, (01)10801902161922(17)241231(10)18F23A0021, (01)10801902161922(17)241231(10)18F23A0060, (01)10801902161922(17)250731(10)18F23G0068, (01)10801902161922(17)250930(10)18F23K0012, (01)10801902161922(17)251031(10)18F23K0053, (01)10801902161922(17)251031(10)18F23K0063, (01)10801902161922(17)260131(10)18F24B0051, (01)10801902172874(17)240430(10)18F22E0012, (01)10801902172874(17)240630(10)18F22G0011, (01)10801902172874(17)240630(10)18F22G0016, (01)10801902172874(17)250930(10)18F23K0039, (01)10801902172874(17)251031(10)18F23L0010, (01)10801902172874(17)251031(10)18F23L0031, (01)10801902172874(17)251031(10)18F23L0039, (01)10801902172874(17)251130(10)18F23K0008
- Product
- Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
- Recalling firm
- ARROW INTERNATIONAL Inc.
- Quantity
- 3138 units
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.
- Recall date
- June 12, 2024
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)