Medline Sub-G Endotracheal Tube with Subglottic recall
Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5…
Why was Medline Sub-G Endotracheal Tube with Subglottic recalled?
Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI numbers for the individual tubes: (1) DYNJ18860 - UDI-DI 20888277652672; (2) DYNJ18865 - UDI-DI 20888277652689; (3) DYNJ18870 - UDI-DI 20888277652696; (4) DYNJ18875 - UDI-DI 20888277652702; (5) DYNJ18880 - UDI-DI 20888277652719; (6) DYNJ18885 - UDI-DI 20888277657318. Kit code information: (1) Kit ACC010502 - Kit UDI-DI: 10193489846171; Case UDI-DI: 40193489846172; lot numbers: 23FDC233, 23FDB211, 23EDC153, 23EDB775, 23DDB280, 23CDB943, 23CDA556, 22LDA029, 22IDA550, 22HDA813, 22FDC033, 22CDC284, 21JDC054, 21IDB345, 21HDB405, and 21DDB305; (2) Kit ACC010527 - Kit UDI-DI: 10193489884333; Case UDI-DI: 40193489884334; lot numbers: 23JDA443, 23IDA363, 23IDA232, 23HDC331, 23HDA117, 23GDB600, 23FDC372, 23FDB913, 23FDA966, 23FDA310, 23EDC371, 23EDC154, 23EDC094, 23EDB732, 23EDA387, 23CDB838, 23CDA557, 23BDC008, 23BDA945, 23ADB791, 22LDA005, 22JDB800, 22HDB976, 22HDA451, 22GDB041, 21LDB937, 21LDB416, 21JDB262, 21IDB491, 21IDA193, 21GDB207, and 21FDC754; (3) Kit ACC10540 - Kit UDI-DI: 10193489977547; Case UDI-DI 40193489977548; lot number: 23ADB410; (4) Kit ACC10540A - Kit UDI-DI 10195327330088; Case UDI-DI 40195327330089; lot numbers: 23HDA588, 23GDB107, and 23BDA684; (5) Kit ACC010717 - Kit UDI-DI 10195327458997; Case UDI-DI 40195327458998; lot numbers: 24ADA577, 23LDA731, and 23IDA629; (6) Kit DYNDJ1132 - Kit UDI-DI 10195327454821; Case UDI-DI 40195327454822; lot numbers: 23LDA733, 23LDA617, 23JDC021, and 23HDA569; (7) Kit DYNDJ1133 - Kit UDI-DI 10195327428532; Case UDI-DI 40195327428533; lot numbers: 24BMB124, 23LMB682, 23HMF259, and 23HMD765; (8) Kit DYNJ909501 - Kit UDI-DI 10195327384852; Case UDI-DI 40195327384853; lot number: 23GBI504; (9) Kit DYNJ909501A - Kit UDI-DI 10195327549817; Case UDI-DI 40195327549818; lot numbers: 24BBD129, 24ABK281, and 23KBR773; (10) Kit DYNJAA269 - Kit UDI-DI 10195327209049; Case UDI-DI 40195327209040; lot numbers: 24BBE224, 24ABM718, 24ABD835, 23LBB055, 23JBP694, 23JBQ783, 23IBU821, 23IBF348, 23HBP188, 23CBX318, 23CBX319, 23BBR494, 23ABI514,22LBI988, 22KBH009, and 22JBH478.
- Product
- Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5…
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 168,632 tubes and 13,092 kits
- Distribution
- Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.
- Recall date
- June 5, 2024
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)