FDAClass IIMay 29, 2024

Incisive CT - Computed Tomography X-Ray System intended recall

Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143

Why was Incisive CT - Computed Tomography X-Ray System intended recalled?

Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: (01) 00884838085015 (21) + Serial number Serial Numbers- Applied to FCO72800819: 500046, 500064, 500074, 500102, 500110, 500111, 500145, 500168, 500178, 500211 500216, 500217, 500222, 500225, 500230, 500240, 500242. Serial Number-Applied to FCO72800820 : 500615

Product
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143
Recalling firm
Philips North America Llc
Quantity
18 units
Distribution
Worldwide - US Nationwide distribution.
Recall date
May 29, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)