Kelyniam Custom recall
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1002
Why was Kelyniam Custom recalled?
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C. Case Numbers: CSI082114-CT1 CSI081214-TB1 CSI020216-LK1 CSI092016-LD1R CSI100116-BM1A CSI092616-YB1 CSI092616-MM1 CSI112316-YB1 CSI050217-LS1 CSI052217-AF1 CSI062317-GM1 CSI071017-NS1B CSI071417-CM1 CSI102617-DL1 CSI101617-JF1A CSI102617-DL1
- Product
- Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1002
- Recalling firm
- Kelyniam Global, Inc.
- Quantity
- 16
- Distribution
- US nationwide distribution.
- Recall date
- May 23, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)