The ARCHITECT Estradiol Kit Intended to measure recall
The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma.
Why was The ARCHITECT Estradiol Kit Intended to measure recalled?
Interaction of Fulvestrant with the ARCHITECT Estradiol assay, leading to falsely elevated estradiol results in patients treated with Fulvestrant.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product List Number 07K7220 with Lot #: 55900UI01with expiration date of 04/23/2016, Lot #: 55908UI01 with expiration date of 04/29/2016, Lot #: 55941UI00 with expiration date of 04/23/2016, Lot #: 57929UI00 with expiration date of 07/16/2016, Lot #: 57931UI01 with expiration date of 07/27/2016, Lot #: 60105UI00 with expiration date of 10/02/2016 and Lot #: 61106UI00 with expiration date of 11/27/2016. Product List Number 07K7225 - Lot #: 55900UI00 with expiration date of 04/23/2016, Lot #: 55908UI00 with expiration date of 04/29/2016, Lot #: 57929UI01 with expiration date of 07/16/2016, Lot #: 57931UI00 with expiration date of 08/27/2016, Lot #: 60105UI01 with expiration date of 02/10/2016, Lot #: 61106UI01 with expiration date of 11/27/2016 and Lot #: 62263UI00 with expiration date of 11/27/2016.
- Product
- The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma.
- Recalling firm
- Abbott Laboratories
- Quantity
- 39,346 units
- Distribution
- Worldwide Distribution -- US, to the following states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY; and, the countries of ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVI, BRAZIL, BRUNEI, BULGARIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GREECE, GUATEMALA, HONDURAS, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MALDIVES, MALI, MAURITIUS, MEXICO, MONTENEGRO, MOROCCO, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TOGO, TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM and ZIMBABWE.
- Recall date
- June 8, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)