FDAClass IIJune 4, 2025

NxStage PureFlow-B Solution recall

NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.

Why was NxStage PureFlow-B Solution recalled?

Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a. UDI-DI number: M535RFP4000. Device Catalog Number RFP-400. Lot Numbers: Q2407307, Q2407308, Q2407309. b. UDI-DI number: M535RFP4010. Device Catalog Number RFP-401. Lot Numbers: Q2407061, Q2407147, Q2407148, Q2407149, Q2407313, Q2407386, Q2407387, Q2407388, Q2407389, Q2407390. c. UDI-DI number: M535RFP4020. Device Catalog Number RFP-402. Lot Numbers: Q2407213, Q2407214. d. UDI-DI number: M535RFP4030. Device Catalog Number RFP-403. Lot Number: Q2407311. e. UDI-DI number: M535RFP4040. Device Catalog Number RFP-404. Lot Numbers: Q2407062, Q2407063, Q2407064, Q2407065, Q2407067, Q2407068, Q2407069, Q2407070, Q2407071, Q2407217, Q2407218, Q2407219, Q2407220, Q2407221, Q2407222, Q2407223, Q2407224, Q2407225, Q2407301, Q2407302, Q2407303, Q2407304, Q2407305, Q2407306, Q2407391, Q2407392, Q2407405, Q2407406. f. UDI-DI number: M535RFP4060. Device Catalog Number RFP-406. Lot Numbers: Q2407072, Q2407137, Q2407140, Q2407141, Q2407216, Q2407310, Q2407312. g. UDI-DI number: M535RFP4530. Device Catalog Number RFP-453. Lot Number: Q2407066. h. UDI-DI number: M535RFP4540. Device Catalog Number RFP-454. Lot Numbers: Q2407145, Q2407146, Q2407226. j. UDI-DI number: M535RFP4560. Device Catalog Number RFP-456. Lot Numbers: Q2407138, Q2407139, Q2407142, Q2407143, Q2407144, Q2407227.

Product
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
Recalling firm
NxStage Medical Inc
Quantity
115,465 cases (230,930 units)
Distribution
US Nationwide distribution.
Recall date
June 4, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls