Movable Core Wire Guide recall: Sterility Concern
Movable Core Wire Guide (Tefcor), Guidewire (OUS ONLY) Reference Part Number/GPN TMT-35-145 G01677
Why was Movable Core Wire Guide recalled?
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
TMT-35-145 G01677 UDI-DI: 00827002016774 Lot/Expiration Date: 14808818 22-06-2027
- Product
- Movable Core Wire Guide (Tefcor), Guidewire (OUS ONLY) Reference Part Number/GPN TMT-35-145 G01677
- Recalling firm
- Cook Incorporated
- Quantity
- N/A
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Recall date
- October 5, 2022
- Classification
- Class II
- Status
- Ongoing