EZ Slide Battery/line powered hydraulic table with recall
EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating…
Why was EZ Slide Battery/line powered hydraulic table with recalled?
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014
- Product
- EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating…
- Recalling firm
- Skytron, Div. The KMW Group, Inc
- Quantity
- 458
- Distribution
- Nationwide Distribution to AL, AZ, CO, IA, WI, PA, TN, MA, TX, MI, FL, MS, LA, and CA.
- Recall date
- June 8, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)