Achieva 3.0T TX for PET recall
Achieva 3.0T TX for PET Model Number (REF): 781479
Why was Achieva 3.0T TX for PET recalled?
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: N/A Serial Numbers: 43102
- Product
- Achieva 3.0T TX for PET Model Number (REF): 781479
- Recalling firm
- Philips North America Llc
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
- Recall date
- May 29, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)