FDAClass IIJune 25, 2014

G7 Acetabular System Positioning Guide Rod recall

G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.

Why was G7 Acetabular System Positioning Guide Rod recalled?

Guide rod may fracture during impaction of the acetabular shell and be retained in the inserter or the patient and could result in a delay in surgery.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

PN:110003458. Lots 424960, ZB130702, ZB130703, ZB130704, ZB130705, ZB130701, ZB131201, 360877, 377304 and ZB140201.

Product
G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.
Recalling firm
Biomet, Inc.
Quantity
1047 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Netherlands, New Zealand, and Japan.
Recall date
June 25, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)