Allura Xper Release 8.2 recall
Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029…
Why was Allura Xper Release 8.2 recalled?
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Devices with Laird Chiller.
- Product
- Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029…
- Recalling firm
- Philips Electronics North America Corporation
- Quantity
- 6992 (total for all devices)
- Distribution
- Nationwide, worldwide including Canada
- Recall date
- April 26, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)