FDAClass IIMay 6, 2020

Replaceable lead set for the Wireless Acquisition recall

Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following…

Why was Replaceable lead set for the Wireless Acquisition recalled?

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Numbers: 9293-046-60 - 9293-047-71; 9293-017-50 - 9293-017-51; 9293-026-50 - 9293-026-51; 9293-061-50 - 9293-061-53; 9293-033-52 - 9293-033-53 and 9293-034-50.

Product
Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following…
Recalling firm
WELCH ALLYN, INC/MORTARA
Quantity
78,521 units total
Distribution
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Recall date
May 6, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls