Lead diagnostic electrocardiograph under the following recall
Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and…
Why was Lead diagnostic electrocardiograph under the following recalled?
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Numbers: ELI10-XXX-XXXXX, ELI350-XXX-XXXXX, ELI150C-XXX-XXXXX, ELI250C-XXX-XXXXX, ELI230-XXX-XXXXX, ELI280-XXX-XXXXX, ELI380-XYYZZ, RSCRIBE-xxx-xxxxx and CC-XXX-XXXX.
- Product
- Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and…
- Recalling firm
- WELCH ALLYN, INC/MORTARA
- Quantity
- 78,521 units total
- Distribution
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Recall date
- May 6, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)