Navigation System II-Cart recall
Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The…
Why was Navigation System II-Cart recalled?
Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers 100019, 100984, 100965, 100295, 100979, 100773, 100946, 100797, 100947, 100956, 100966, 100649, 100967, 100944, 100941, 100945, 100970, 100950, 100976, 100951, 100728, 100952, 100248, 100953, 100641, 100954, 100463, 100955, 100981, 100957, 100978, 100958, 100106, 100959, 100939, 100960, 100973, 100961, 100236, 100962, 100237, 100963, 100417, 100964. 100940, 100968, 100975, 100969, 100977, 100972, 100105, 100974, 100214, 100980, 100492, 100982, 100493, 100983, 100948, 100986, 100704, 100987, 100949, 100988, 100227, 100989, 100971, 100133, 100134, 100942.
- Product
- Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The…
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Quantity
- 72
- Distribution
- Worldwide Distribution - US Nationwide in the states of AR, CO, FL, GA, IN, KS, KY, MA, MI, NC, NE, NJ, NY, OH, OK, PA, TX, WA, WY and the countries of Canada, Europe, and Asia.
- Recall date
- June 24, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)