FDAClass IIMay 6, 2020

Philips Azurion 7 M20 -XperGuide Software hosted in recall: Software Issue

Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic…

Why was Philips Azurion 7 M20 -XperGuide Software hosted in recalled?

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software:Philips Azurion 2.0 Interventional Fluoroscopic X-ray system with Interventional Workspot 1.5 US Serial Numbers: 637, 200, 909, 1276, 1309, 987 Canada: 1016 ROW: 11169 1129 1143 971 1135 824 934 1089 1008 1165 1080 1019 1248 523 1172 Expanded Recall OUS: 1223, 1330, 703424

Product
Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic…
Recalling firm
Philips North America, LLC
Quantity
22 units (6 US and 16 OUS); Expanded 3units; Total: 25
Distribution
Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
Recall date
May 6, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls