FDAClass IIMay 6, 2020

Obalon Balloon System with recall

Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used…

Why was Obalon Balloon System with recalled?

During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ObalonR Balloon System with Navigation and Touch. UDI: 00859810006104; Model No.: Model 4300 Touch Dispenser;Software version 02.00.00.0076. Serial Numbers: 0097, 0076, 0077, 0078, 0079, 0080, 0082, 0083, 0084, 0085, 0086, 0087 0088, 0089, 0090, 0091, 0092, 0093, 0094, 0095, 0097, 0099, 0100, 0101, 0102, 0103, 0104, 0106, 0107, 0109, 0111, 0112, 0113, 0115, 0118, 0119, 0120, 0121, 0122, 0125, 0126, 0127, 0129, 0130, 0133.

Product
Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used…
Recalling firm
Obalon Therapeutics Inc
Quantity
45 units
Distribution
US Nationwide distribution including the states of VA, NY, GA, TX, TN, OR, FL, LA, CT, and NJ.
Recall date
May 6, 2020
Classification
Class II
Status
Completed
Agency
FDA (device)
Open FDA enforcement record Latest recalls